Milwaukee MW600 PRO Portable Dissolved Oxygen Meter
search
  • Milwaukee MW600 PRO Portable Dissolved Oxygen Meter

Milwaukee MW600 PRO Portable Dissolved Oxygen Meter

$425.99

The MW600 is a Portable Dissolved Oxygen meter ideal for use in school laboratories.

Quantity
Out-of-Stock

Dissolved Oxygen measurements are also very important in fish farms and ponds, where Oxygen levels are continuously monitored to obtain optimal reproduction.

The MW600 calibrates easily in 2 points (at 100% saturated air and in 0 Oxygen solution) and has Automatic Temperature Compensation which guarantees the highest accuracy. The low battery warning, easy to replace screw on cap membranes make this meter very simple to operate. Rugged Carrying Case (Optional) provides handy on-site meter calibration and measurements.

MW600 is supplied complete with a MA840 D.O. polarographic probe with 3 m cable, calibration screwdriver, 2 spare membranes, MA7040 (20 mL) electrolyte solution, battery and instructions.

DESIGN FEATURES

  • Fast, easy-to-read results with large digital LCD readout
  • Repeatable, accuracy to ±1.5% FS
  • Lightweight and portable
  • Simple 2 point open air calibration (see video below)
  • Uses MA840 reinforced polarographic probe with an oxygen-permeable polytetrafluroethylene (PTFE) membrane sensor
  • 5 o-rings, membranes and one protective cap included
  • 10 feet probe cable for extra flexibility when testing
  • 70 hours of battery life with continuous use (battery included) with low battery warning

CARE AND USE

“OPEN AIR” CALIBRATION PROCEDURE

1. Remove the membrane

Take the membrane off the end of the probe and dry the diode off.

 

2. Zero out the display

Connect the probe to the base unit and turn the unit on. Your display should read 0.0. If it does not, go to the position “0” screw on the front of the unit with the screw driver provided and through the access hole make the unit display read 0.0.

 

3. Fill the membrane

Go to the membrane and put the “O” ring inside the membrane on the shoulder at the base of the threads. Fill the membrane to just short of the “O” ring with MA9071 electrolyte solution. Be sure that there are no bubbles inside the membrane. (Air bubbles are the main cause of calibration failure)

 

4. Attach the membrane

Screw the membrane on to the probe.

 

5. Let stablize

Turn the meter on and hang the probe off the end of a table pointing down and let the unit display stabilize (this may take 5 to 15 minutes)


6. Adjust

Push down and hold the “Cal” button on the front face of the meter and with the screwdriver go to the front Cal position “100%” and make the display read 100.

 

You are now calibrated.

SPECIFICATIONS

  • O2 Range: 0.0 to 19.9 mg/L
  • O2 resolution: 0.1 mg/L
  • O2 Accuracy: ±1.5% Full Scale
  • Calibration: Manual on 2 points (zero and slope)
  • Temperature Compensation: Automatic, from 0 to 30°C
  • Probe: MA840 refillable polarographic probe included
  • Membranes: 5 included
  • Environment: 0 to 50°C / 32 to 122°F; max RH 95%
  • Battery Type: 1 x 9V alkaline (included)
  • Battery Life: Approx. 70 hours of continuous use
  • Dimensions: 5.7 x 3.1 x 1.6 inches
  • Weight: 7.5 ounces with battery

 

MW600
Comments (0)
No customer reviews for the moment.